STIS can be used to obtain images in undispersed light in the optical and ultraviolet. When STIS is used in imaging mode, the appropriate clear or filtered aperture on the slit wheel is rotated into position, and a mirror on the Mode-Selection Mechanism is moved into position (see Figure 3.1:).
Table 5.1 provides a complete summary of the clear and filtered apertures available for imaging with each detector. In Figure 5.6 through Figure 5.9 we show the integrated system throughputs.
Aperture Name |
Filter |
Central Wavelength ( ![]() |
FWHM ( ![]() ![]() |
Field of View (arcsec) |
|
Detector |
ref. page |
---|---|---|---|---|---|---|---|
Visible - plate scale ![]() |
|||||||
50CCD |
Clear |
5850 |
4410 |
52 x 52 |
|
STIS/CCD |
|
F28X50LP |
Optical longpass |
7230 |
2720 |
28 x 521 |
|
STIS/CCD |
|
F28X50OIII |
[O III] |
5007 |
5 |
28 x 52 1 |
|
STIS/CCD |
|
F28X50OII |
[O II] |
3740 |
80 |
28 x 52 1 |
|
STIS/CCD |
|
50CORON |
Clear + coronagraphic fingers |
5850 |
4410 |
52 x 52 |
|
STIS/CCD |
|
Ultraviolet - plate scale ~0.0246 arcseconds per pixel2 |
|||||||
25MAMA |
Clear |
2220 1370 |
1200 320 |
25 x 25 |
|
STIS/NUV-MAMA STIS/FUV-MAMA |
|
F25QTZ |
UV near longpass |
2320 1590 |
1010 220 |
25 x 25 |
|
STIS/NUV-MAMA STIS/FUV-MAMA |
|
F25SRF2 |
UV far longpass |
2270 1480 |
1110 280 |
25 x 25 |
|
STIS/NUV-MAMA STIS/FUV-MAMA |
|
F25MGII |
Mg II |
2800 |
70 |
25 x 25 |
|
STIS/NUV-MAMA |
|
F25CN270 |
Continuum near 2700 Å |
2700 |
350 |
25 x 25 |
|
STIS/NUV-MAMA |
|
F25CIII |
C III] |
1909 |
70 |
25 x 25 |
|
STIS/NUV-MAMA |
|
F25CN182 |
Continuum near 1800 Å |
1820 |
350 |
25 x 25 |
|
STIS/NUV-MAMA |
|
F25LYA |
Lyman- ![]() |
1216 |
85 |
25 x 25 |
|
STIS/FUV-MAMA |
|
Neutral-Density-Filtered Imaging |
|||||||
F25NDQ1 F25NDQ2 F25NDQ3 F25NDQ4 |
ND=10-1 ND=10-2 ND=10-3 ND=10-4 |
1150-10,300 Å |
13.4 x 9.7 13.8 x 15.1 11.4 x 15.3 11.8 x 9.5 |
|
STIS/NUV-MAMA STIS/FUV-MAMA STIS/CCD |
||
F25ND3 |
Neutral-density filter, ND=10-3 |
1150-10,300 Å |
25 x 25 |
|
STIS/NUV-MAMA STIS/FUV-MAMA STIS/CCD |
||
F25ND5 |
ND=10-5 |
1150-10,300 Å |
25 x 25 |
|
STIS/NUV-MAMA STIS/FUV-MAMA STIs/CCD |
1The dimensions are 28 arcsec on AXIS2 =Y and 52 arcsec on AXIS1 =X. See Figure 3.2 and Figure 11.1:.2The MAMA plate scales differ by about 1% in the AXIS1 and AXIS2 directions, a factor that must be taken into account when trying to add together rotated images. Also, the FUV-MAMA uses a different mirror in the filtered and unfiltered modes. In the filtered mode, the plate scale is 0.3% larger (more arcsec/pixel). Information on geometric distortions can be found in Chapter 14. |
The Advanced Camera for Surveys (ACS) was installed on HST during servicing mission SM3B in March of 2002, and has been performing very well since that time. Compared with STIS, ACS offers detectors with a much larger field of view, significantly higher throughputs at most wavelengths, a wider selection of filters, better suppression of point spread function (PSF) wings, and newer CCD detectors with far less accumulated radiation damage. Observers will therefore find that, especially at optical wavelengths, most imaging programs are better and more efficiently done with ACS than with STIS. There will, however, still be some cases where imaging with STIS is a better choice. The following points should be considered when choosing between STIS with ACS imaging:
F28X50OII
filter (3727 Å), and the STIS NUV-MAMA F25MGII
(2800 Å) and F25CIII
(1900 Å) filters. The NUV-MAMA intermediate band F25CN182
and F25CN270
filters may also prove useful for some programs.
F606W
or F814W
will usually be a better choice than the STIS 50CCD
or F28X50LP
configurations (see
Figure 5.1:)
.
F220W
or F250W
filter should be considered as an alternative to broadband imaging with the STIS NUV-MAMA (see Figure 5.2 and Figure 5.3:). The NUV MAMA has no read noise, a very low red sensitivity (so that filter redleaks have very little impact), and a lower dark current (the MAMA's dark current advantage over the HRC should increase as the HRC CCD ages), while the HRC has a higher peak sensitivity and, unlike the NUV-MAMA detector, is not subject to bright object constraints. The NUV-MAMA does have a slightly smaller pixel size than the HRC, but the much cleaner PSF of the HRC will often offset this advantage when attempting to detect faint sources near bright objects. Furthermore, there is a significant focus change with position across the STIS NUV-MAMA (see Section 5.1.3). The final choice between the NUV-MAMA and the ACS HRC will depend on the details of each program's science requirements.
TIME-TAG
mode.
The ACS Instrument Handbook for Cycle 13 provides additional comparisons between the capabilities of STIS and ACS. Ultimately, observers should use the exposure time calculators for STIS and ACS to decide which of the two instruments is better suited for their science.
Figure 5.1: The total system throughputs of the STIS50CCD
and F28X50LP
configurations are compared to the throughputs of several broad band ACS WFC filters.25MAMA
, F25QTZ
, and F25SRF2
apertures is compared with that of the ACS HRC F220W
and F250W
.
While ACS has clear advantages over either STIS or WFPC2 for most prime GO imaging science, the great success of ACS also means that there will be ample opportunities for parallel observing with the other HST instruments. It is therefore still useful to compare the imaging capabilities of STIS and WPFC2.
50CCD
aperture have significantly higher sensitivity to faint sources than WFPC2 (figure 5.6 on page 63). The CCD long-pass imaging filter, F28X50LP
, has similar sensitivity longward of 5500 Å. The STIS CCD also has a low read noise and dark current; thus STIS CCD observations with the clear or long-pass apertures are extremely sensitive to faint sources (see figure 5.6 on page 63). The STIS CCD clear imaging mode is especially useful when no color information is needed, for example, for finding faint variable sources, or imaging the faintest possible sources in a given integration time.
There are several important points about imaging with STIS which should be kept in mind:
F28X50OIII
and F25MGII
) have substantial red leaks (see Figure 5.11 and Figure 5.16, respectively).
GAIN=1
; programs at high count levels should use GAIN=4
. At GAIN=4
the CCD exhibits a modest read-noise pattern that is correlated on scales of tens of pixels. (See Section 7.1.10.)
These caveats are not intended to discourage observers from using STIS for imaging; indeed, for many imaging projects, particularly those not requiring a large field of view or the range of filters provided by WFPC2 and ACS, STIS may be the best choice.
In Figure 5.6: above, we show the throughput (where the throughput is defined as the end-to-end effective area divided by the geometric area of a filled, unobstructed, 2.4 meter aperture) of the three STIS clear imaging modes, with the CCD, the NUV-MAMA, and the FUV-MAMA. Superposed on this plot, we show the broadband WFPC2 throughputs. In Figure 5.7, Figure 5.8, and Figure 5.9, we show the throughputs of the full set of available filters for the CCD, the NUV-MAMA, and the FUV-MAMA, respectively.
Figure 5.7: STIS's CCD Clear and Filtered Imaging Mode Throughputs
In Table 5.2 below, we give the A0 V star V magnitude reached during a one-hour integration which produces a signal-to-noise ratio of 10 integrated over the number of pixels needed to encircle 80% of the PSF flux. The observations are assumed to take place in shadow
, with average zodiacal background. These examples are for illustrative purposes only. Since sky background becomes important for the source count rates in Table 5.2, the exposure times will be much longer outside SHADOW
. For instance, if a 27.7 mag A star were observed in Bright Earth, the exposure time to reach signal-to-noise=10 with CCD clear would be about 4 times longer.
In Chapter 14 we present, for each imaging mode, plots of exposure time versus magnitude to achieve a desired signal-to-noise ratio. These plots, which are referenced in the individual imaging-mode sections below, are useful for getting an idea of the exposure time you need to accomplish your scientific objectives. More detailed estimates can be made either by using the sensitivities given in Chapter 14 or by using the STIS Imaging Exposure Time Calculator
. The exposure time calculator is also available as part of the APT
package.
Both CCD and MAMA imaging observations are subject to saturation at high total accumulated counts per pixel: the CCD due to the depth of the full well and the saturation limit of the gain amplifier for CCDGAIN = 1
; and the MAMA due to the 16-bit format of the buffer memory (see Section 7.2.1 and Section 7.4.1). In Chapter 14, saturation levels as functions of source magnitude and exposure time are presented in the S/N plots for each imaging mode.
Space Telescope Science Institute http://www.stsci.edu Voice: (410) 338-1082 help@stsci.edu |