STIS can be used to obtain images in undispersed light in the optical and ultraviolet. When STIS is used in imaging mode, the appropriate clear or filtered aperture on the slit wheel is rotated into position, and a mirror on the Mode-Selection Mechanism is moved into position (see Figure 3.1).
Table 5.1 provides a complete summary of the clear and filtered apertures available for imaging with each detector. In Figure 5.1 through Figure 5.4 we show the integrated system throughputs.
Aperture Name |
Filter |
Central Wavelength (c in Å) |
FWHM ( in Å) |
Field of View (arcsec) |
|
Detector |
ref. page |
---|---|---|---|---|---|---|---|
Visible - plate scale
per pixel |
|||||||
50CCD |
Clear |
5850 |
4410 |
52 x 52 |
|
STIS/CCD |
|
F28X50LP |
Optical longpass |
7230 |
2720 |
28 x 521 |
|
STIS/CCD |
|
F28X50OIII |
[O III] |
5007 |
5 |
28 x 52 1 |
|
STIS/CCD |
|
F28X50OII |
[O II] |
3740 |
80 |
28 x 52 1 |
|
STIS/CCD |
|
50CORON |
Clear + coronographic fingers |
5850 |
4410 |
52 x 52 |
|
STIS/CCD |
|
Ultraviolet - plate scale ~0.0246 arcseconds per pixel2 |
|||||||
25MAMA |
Clear |
2220 1370 |
1200 320 |
25 x 25 |
|
STIS/NUV-MAMA STIS/FUV-MAMA |
|
F25QTZ |
UV near longpass |
2320 1590 |
1010 220 |
25 x 25 |
|
STIS/NUV-MAMA STIS/FUV-MAMA |
|
F25SRF2 |
UV far longpass |
2270 1480 |
1110 280 |
25 x 25 |
|
STIS/NUV-MAMA STIS/FUV-MAMA |
|
F25MGII |
Mg II |
2800 |
70 |
25 x 25 |
|
STIS/NUV-MAMA |
|
F25CN270 |
Continuum near 2700 Å |
2700 |
350 |
25 x 25 |
|
STIS/NUV-MAMA |
|
F25CIII |
C III] |
1909 |
70 |
25 x 25 |
|
STIS/NUV-MAMA |
|
F25CN182 |
Continuum near 1800 Å |
1820 |
350 |
25 x 25 |
|
STIS/NUV-MAMA |
|
F25LYA |
Lyman- |
1216 |
85 |
25 x 25 |
|
STIS/FUV-MAMA |
|
Neutral-Density-Filtered Imaging |
|||||||
F25NDQ1 F25NDQ2 F25NDQ3 F25NDQ4 |
ND=10-1 ND=10-2 ND=10-3 ND=10-4 |
1150-10,300 Å |
12 x 12 12 x 12 12 x 12 12 x 12 |
|
STIS/CCD |
||
F25ND3 |
Neutral-density filter, ND=10-3 |
1150-10,300 Å |
25 x 25 |
|
STIS/NUV-MAMA STIS/FUV-MAMA |
||
F25ND5 |
ND=10-5 |
1150-10,300 Å |
25 x 25 |
|
STIS/NUV-MAMA STIS/FUV-MAMA |
1 The dimensions are 28 arcsec on AXIS2=Y and 52 arcsec on AXIS1=X. See Figure 3.2 and Figure 11.1 on page 213. 2 The MAMA plate scales differ by about 1% in the AXIS1 and AXIS2 directions, a factor that must be taken into account when trying to add together rotated images. Also, the FUV-MAMA uses a different mirror in the filtered and unfiltered modes. In the filtered mode, the plate scale is 0.3% larger (more arcsec/pixel). Further information on geometric distortions can be found in Chapter 14 |
Despite its limited complement of filters, STIS brings several valuable imaging capabilities to HST:
F28X50LP
, has similar sensitivity longward of 5500 Å.The STIS CCD also has a low read noise and dark current; thus STIS CCD observations with the clear or longpass apertures are extremely sensitive to faint sources (see Figure 5.6). The STIS CCD clear imaging mode is especially useful when no color information is needed, for example, for finding faint variable sources, or imaging the faintest possible sources in a given integration time.
TIME-TAG
mode.
There are several important points about imaging with STIS which should be kept in mind:
F28X50OIII
and F25MGII
) have substantial red leaks (see Figure 5.7 and Figure 5.14, respectively).
GAIN=1
; programs at high count levels should use GAIN=4
. At GAIN=4
the CCD exhibits a modest read-noise pattern that is correlated on scales of tens of pixels. (See Analog-to-Digital Conversion: Selecting the CCDGAIN.)
These caveats are not intended to discourage observers from using STIS for imaging; indeed, for many imaging projects, particularly those not requiring a large field of view or range of filters provided by WFPC2 and ACS, STIS may be the best choice.
ACS will be installed in a future Servicing Mission. ACS is designed primarily as an imager, whereas the main scientific capabilities of STIS are spectroscopic. Therefore observers will find that many imaging programs are more efficiently done with ACS than with STIS. This applies to the optical wavelength range in particular, where STIS offers only a limited choice of filters. Together with its larger field of view, this will make ACS the instrument of choice for imaging at optical wavelengths in most cases.
The choice between STIS and ACS for ultraviolet imaging largely depends on the details of the science requirements. ACS offers a larger variety of filters. On the other hand, STIS has better sampling of the PSF, thus providing higher spatial resolution, and it has a much lower dark current. Observers must critically examine how the different field of view of STIS and ACS affect their science. ACS with its larger field of view will be the preferred instrument for survey-type studies. For high-resolution imaging of individual objects whose size is small in comparison with the detector, bright object considerations may become important. Remember that it is the observer's responsibility to provide UV flux information of all field objects that happen to fall into the field of view. The larger the field size, the higher the likelihood of including such field objects.
The ACS Instrument Handbook for Cycle 11 provides additional comparisons between the capabilities of STIS and ACS. Ultimately, observers should use the exposure time calculators of STIS and ACS to decide which of the two instruments is better suited for their science.
In Figure 5.1 below, we show the throughput (where the throughput is defined as the end-to-end effective area divided by the geometric area of a filled (unobstructed) 2.4 meter aperture) of the three STIS clear imaging modes, with the CCD, the NUV-MAMA, and the FUV-MAMA. Superposed on this plot, we show the broadband WFPC2 throughputs. In Figure 5.2, Figure 5.3, and Figure 5.4, we show the throughputs of the full set of available filters for the CCD, the NUV-MAMA, and the FUV-MAMA, respectively.
We are currently revising the throughput for the CCD imaging modes at wavelengths longer than 6000 Å. Our current estimates suggest that the throughput of the 50CCD
imaging mode should be decreased by roughly 10% near 6000 Å and by as much as 35% near 10,000 Å. The ratio of the F28X50LP
throughput to the 50CCD
throughput will be increased by about 20%. The exact wavelength dependence of these changes is still under review and the adopted changes will be reported in the HST Spectroscopy STScI Analysis Newsletter (STAN) and on the STScI website.
In Table 5.2 below, we give the A0 V star V magnitude reached during a one-hour integration which produces a signal-to-noise ratio of 10 integrated over the number of pixels needed to encircle 80% of the PSF flux. The observations are assumed to take place in shadow
, with average zodiacal background. These examples are for illustrative purposes only. Since sky background becomes important for the source count rates in Table 5.2, the exposure times will be much longer outside SHADOW
. For instance, if a 27.7 mag A star were observed in Bright Earth, the exposure time to reach signal-to-noise=10 with CCD clear would be about 4 times longer.
In Chapter 14 we present, for each imaging mode, plots of exposure time versus magnitude to achieve a desired signal-to-noise ratio. These plots, which are referenced in the individual imaging-mode sections below, are useful for getting an idea of the exposure time you need to accomplish your scientific objectives. More detailed estimates can be made either by using the sensitivities given in Chapter 14 or by using the STIS Imaging Exposure Time Calculator
. A web based version of this tool can be reached from the STIS web page (http://www.stsci.edu/instruments/stis/
), under Tools.
Both CCD and MAMA imaging observations are subject to saturation at high total accumulated counts per pixel: the CCD due to the depth of the full well and the saturation limit of the gain amplifier for CCDGAIN = 1
; and the MAMA due to the 16-bit format of the buffer memory (see CCD Saturation and MAMA Saturation-Overflowing the 16 Bit Buffer). In Chapter 14, saturation levels as functions of source magnitude and exposure time are presented in the S/N plots for each imaging mode.
Space Telescope Science Institute http://www.stsci.edu Voice: (410) 338-1082 help@stsci.edu |