25MAMA-FUV-MAMA Clear
Description
The 25MAMA
mode with the FUV-MAMA detector provides high-throughput broad-band far-UV imaging with the highest available throughput at ~1500 Å and the highest possible spatial resolution. The field of view is 25 x 25 arcsec.
Recommended Uses
Recommended for broad-band near-UV imaging of sources where sky background is not a consideration.
Special Considerations
For long exposures of faint targets, sky background is likely to be a limiting factor. Observers should consider the use of DARKTIME
or (preferably) use the F25QTZ
filter, which has nearly the same throughput but rejects geocoronal Lyman- and OI 1302 Å air glow.
Pivot (Å) |
FWHM (Å) |
AB mag zeropoint |
Speak |
B |
R80 (arcsec) |
Flux in central pixel |
1374.3 |
324.1 |
23.595 |
1.30e+14 |
366.9 |
0.32 |
8% |
Figure 14.66: 25MAMA FUV-MAMA Integrated System Throughput and Redleak
Table 14.26: Throughput and Sensitivity for 25MAMA FUV-MAMA
l |
Sensitivity |
Throughput % |
1150. |
1.72E13 |
0.66 |
1175. |
5.41E13 |
2.02 |
1200. |
9.51E13 |
3.48 |
1225. |
1.25E14 |
4.48 |
1250. |
1.29E14 |
4.54 |
1300. |
1.25E14 |
4.23 |
1350. |
1.18E14 |
3.85 |
1375. |
1.14E14 |
3.63 |
1400. |
1.08E14 |
3.39 |
1425. |
9.99E13 |
3.08 |
1450. |
9.14E13 |
2.77 |
1475. |
8.28E13 |
2.46 |
1500. |
7.41E13 |
2.17 |
1550. |
5.87E13 |
1.66 |
1600. |
4.43E13 |
1.21 |
1650. |
3.34E13 |
0.89 |
1700. |
2.36E13 |
0.61 |
1750. |
1.67E13 |
0.42 |
1800. |
1.16E13 |
0.28 |
1850. |
4.45E12 |
0.11 |
1875. |
2.17E12 |
0.05 |
Figure 14.67: Point Source S/N vs. V+AB for the 25MAMA FUV-MAMA mode.
Figure 14.68: Extended Source S/N vs. V+AB for the 25MAMA FUV-MAMA mode. Top curves are for an area of 0.2 arcsec2; bottom curves are for 1 arcsec2. Average sky assumed.
Table 14.27: Radial Profile for 25MAMA
Radius |
|
Fraction of Energy Encircled |
Pixels |
Arcsec |
Intensity |
1 |
0.024 |
3.267e-02 |
0.141 |
2 |
0.037 |
2.107e-02 |
0.236 |
2 |
0.049 |
1.345e-02 |
0.331 |
3 |
0.073 |
6.141e-03 |
0.467 |
4 |
0.098 |
2.981e-03 |
0.557 |
5 |
0.122 |
1.485e-03 |
0.614 |
10 |
0.245 |
3.377e-04 |
0.746 |
15 |
0.367 |
1.676e-04 |
0.834 |
20 |
0.490 |
9.542e-05 |
0.900 |
25 |
0.613 |
4.543e-05 |
0.947 |
30 |
0.735 |
2.065e-05 |
0.972 |
40 |
0.980 |
5.088e-06 |
0.994 |
Figure 14.69: Point Source Encircled Energy for 25MAMA FUV-MAMA