The STIS PRISM
is used with the STIS/NUV-MAMA
and provides spectra from 1150 to 3620 Å at resolving powers of ~500 in the ultraviolet declining to ~10 in the optical. A highly idealized schematic of a STIS objective-prism observation of a star cluster is shown in Figure 4.10. This example illustrates the power of the prism mode to simultaneously provide spectra covering a wide wavelength range of many objects in a single field of view.
The PRISM
can be used at two wavelength settings, 1200 and 2125 Å. These are approximately the wavelengths that will lie at the center of AXIS1
on the detector for the two settings. The relationship between wavelength and pixel number along the central spectral trace is shown in Figure 4.11 for each setting. The dispersion as a function of wavelength is shown in Figure 4.12 for each setting.
PRISM
at wavelength settings 1200 and 2125 Å.
The prism
can be used with the clear MAMA aperture (25MAMA
) or with either longpass ultraviolet filtered aperture (F25SRF2
or F25QTZ
) to provide a 25 x 25 arcsecond field of view (see also F25SRF2-NUV-MAMA, Longpass and F25ND5-NUV-MAMA). The longpass filter F25SRF2
blocks geocoronal Lyman- 1216 Å and the F25QTZ
longpass filter blocks both geocoronal Lyman- and geocoronal O I 1302 Å, significantly reducing the background from these lines (which is otherwise spread throughout the image) at the price of losing the short-wavelength range of the spectrum. In addition, the neutral-density filters ( Table 5.1) are supported for PRISM
spectroscopy, as are the 52X0.05, 52X0.1, 52X0.2, 52X0.5,
and 52X2
long slits.
Observers will generally want to obtain a direct image of the field they are taking a prism image of, so they can later determine the centering of the objects in their prism data. Because the PRISM
and the mirrors used for imaging are both in the Mode Selection Mechanism, zero-point shifts will occur between the PRISM
and imaging data (see Slit and Grating Wheels). For a discussion of the observations needed to measure these shifts, see Slitless First-Order Spectroscopy.
Note that PRISM
spectroscopy produces images in which, a priori, the wavelength at a given pixel is not known, and source-dependent overlap of spectra can occur. For these reasons, PRISM
spectroscopic data will not be calibrated automatically by the STScI pipeline. Instead, users will have to reduce and analyze their data off-line.
Space Telescope Science Institute http://www.stsci.edu Voice: (410) 338-1082 help@stsci.edu |