There are 15 spectroscopic modes which are summarized in Table 4.1 below. They comprise low and intermediate-resolution first-order modes designed to be used with a complement of long slits over the entire wavelength range, and intermediate and high-resolution echelle modes which have been optimized for point-source observations through short echelle slits and are available only in the ultraviolet (see Figure 4.1).
|
Spectral Range (Å) |
|
Spectral Resolution |
|
|
|
|
|
||
---|---|---|---|---|---|---|---|---|---|---|
Grating |
Complete |
Per Tilt |
|
Scale (Å per pixel) |
Resolving Power (/2) |
No. Prime Tilts1 |
Detector |
Reference material page |
|
Slits |
CCD First-Order Spectroscopy |
|
|
||||||||
G750L |
5240-10,270 |
5030 |
|
4.92 |
530-1040 |
1 |
CCD |
|
52X0.05 52X0.5 52X2 52X0.2F1 |
|
G750M |
5450-10,140 |
570 |
|
0.56 |
4870-9050 |
9 |
CCD |
|||
G430L |
2900-5700 |
2800 |
|
2.73 |
530-1040 |
1 |
CCD |
|||
G430M |
3020-5610 |
286 |
|
0.28 |
5390-10,020 |
10 |
CCD |
|||
G230LB |
1680-3060 |
1380 |
|
1.35 |
620-1130 |
1 |
CCD |
|||
G230MB |
1640-3190 |
155 |
|
0.15 |
5470-10,630 |
11 |
CCD |
|||
MAMA First-Order Spectroscopy |
||||||||||
G230L |
1570-3180 |
1610 |
|
1.58 |
500-1010 |
1 |
NUV-MAMA |
|||
G230M |
1640-3100 |
90 |
|
0.09 |
9110-17,220 |
18 |
NUV-MAMA |
|||
G140L |
1150-1730 |
610 |
|
0.60 |
960-1440 |
1 |
FUV-MAMA |
|||
G140M |
1140-1740 |
55 |
|
0.05 |
11,400-17,400 |
12 |
FUV-MAMA |
|||
MAMA Echelle Spectroscopy |
||||||||||
E230M |
1570-3110 |
800 |
|
/60,000 |
30,000 |
2 |
NUV-MAMA |
|
0.2X0.2,0.2X0.06 |
|
E230H |
1620-3150 |
267 |
|
/228,000 |
114,000 |
6 |
NUV-MAMA |
|
0.2X0.2,0.2X0.09 |
|
E140M |
1150-1710 |
620 |
|
/91,700 |
45,800 |
1 |
FUV-MAMA |
|
0.2X0.2,0.2X0.06 |
|
E140H |
1150-1700 |
210 |
|
/228,000 |
114,000 |
3 |
FUV-MAMA |
|
0.2X0.2,0.2X0.09 |
|
MAMA Prism Spectroscopy |
||||||||||
PRISM |
1150-3620 |
2470 |
|
0.2 - 72 |
2900 - 25 |
1 |
NUV-MAMA |
|
52X0.05,52x0.1, |
1 Number of exposures at distinct tilts needed to cover spectral range of grating with 10% wavelength overlap between adjacent settings. 2 Naming convention gives dimensions of slit in arcseconds. For example, 52X0.1 indicates the slit is 52 arcsec long perpendicular to the dispersion direction and 0.1 arcsec wide in the dispersion direction. The F (e.g., in 52X0.2F1) indicates that it is the fiducial bar to be used for coronographic spectroscopy. F25MGII is supported with all NUV-MAMA gratings and the prism.3 For the MAMA first-order modes, only ~ 25 arcseconds of the long slit projects on the detector. (See also Slits for First-Order Spectroscopy.). 4 Full-aperture clear (50CCD or 25MAMA), longpass-filtered (F25QTZ or F25SRF2 in UV), and neutral-density-filtered slitless spectroscopy are also supported with the appropriate first-order and echelle gratings, as well as the prism. 5 The 6X0.2 and 52X0.05 long slits are also supported for use with the echelle gratings, but with order overlap. The high S/N multislits 0.2X0.2FP(A-E) and 0.2X0.06FP(A-E) (see Chapter 12), as well as the very narrow 0.1X0.03 for maximum spectral resolution, are likewise supported with the echelles. For the echelles, the filtered slits 0.2X0.05ND and 0.3X0.05ND are also supported. The 0.1X0.09 and 0.1X0.2 slits are supported for E230H only. |
To illustrate the broad wavelength coverage provided by STIS, and the relative throughputs achievable across STIS's wavelength regime, we show in Figure 4.2, the system throughput of the four low-resolution, first-order modes on a single plot (where the throughput is defined as the end-to-end effective area divided by the geometric area of a filled, unobstructed, 2.4 meter aperture). To allow you to judge the relative throughputs of different spectroscopic configurations, we plot in Figure 4.3 the efficiency of all grating modes for each of the four primary wavelength regimes on a common plot. These plots allow you to gauge the relative efficiencies of STIS in different configurations. Throughput changes in the first-order modes, determined from monitoring observations since STIS was installed, are discussed in Chapter 13.
Figure 4.2: System Throughput of STIS's Low-Resolution, First-Order Grating Modes
In Table 4.2 below, we give the A0 V type star V magnitude reached during a 1 hour integration which produces a signal-to-noise ratio of 10 in the continuum per spectral resolution element around the peak of the grating response, where we have integrated over the PSF in the direction perpendicular to the dispersion, and assumed the 52X0.2
slit for the first-order gratings and the 0.2X0.2
slit for the echelles.
Both CCD and MAMA observations are subject to saturation at high total accumulated counts per pixel. The CCD can be saturated due to the saturation of the detector itself or of the gain amplifier for CCDGAIN = 1
. MAMA saturation can occur due to the 16-bit format of its memory buffer. The nature of the saturation for CCD and MAMA spectroscopic observations is described in CCD Saturation and MAMA Saturation-Overflowing the 16 Bit Buffer, respectively.
The MAMA detectors are subject to:
We direct MAMA observers to the discussion presented in MAMA Bright-Object Limits. For summary tables of bright-object screening magnitudes for all spectroscopic modes, see MAMA Spectroscopic Bright-Object Limits. It is the observers' responsibility to be sure that proposed observations do not exceed the MAMA bright-object limits.
For the intermediate-resolution gratings and echelles (except E140M), only a portion of the full spectral range of the grating falls on the detector in any one exposure, and the gratings must be scanned (tilted) with a separate exposure taken at each tilt position, in order to cover the full spectral range (see Figure 4.4 and Figure 4.5 below). Accordingly, for these scanned gratings, the user may select a single exposure at a given wavelength, or a series of exposures at different wavelengths to cover a larger wavelength range. The user must choose either prime or secondary settings. The prime settings cover the full spectral range with 10% wavelength overlap between observations taken at adjacent settings. The secondary settings cover selected absorption or emission lines and may be more convenient to use in some applications. We expect the photometric and wavelength calibration accuracies to be higher for the prime settings than for most of the intermediate settings, as calibrations for the latter will be inferred from those taken at prime settings. A few frequently used intermediate settings are being calibrated directly as noted in Chapter 13. The central wavelengths, and corresponding minimum and maximum wavelengths, are presented in the individual grating sections in Chapter 13.
Figure 4.4: Scanned First-Order GratingsIn the near UV, where the CCD has comparable sensitivity to the NUV-MAMA, you may want to consider using the G230LB or G230MB gratings with the CCD instead of the more standard G230L and G230M gratings with the MAMA. You will get improved throughput down to at least 2500 Å, a larger slit length, and use of the CCD rather than the MAMA (see Figure 4.3 and Chapter 13). On the other hand, the CCD has read noise, cosmic-ray sensitivity, hot pixels, and charge transfer efficiency losses. Also, for red objects, scattered light can be more of a problem with the red-sensitive CCD than with the solar-insensitive NUV-MAMA. For a solar-type spectrum, CCD data at wavelengths shorter than 2100 Å are dominated by scattered light.
Space Telescope Science Institute http://www.stsci.edu Voice: (410) 338-1082 help@stsci.edu |