Space Telescope Science Institute  Chapter 4:  Cycle 13 Scientific Instruments  4.2 Advanced Camera for Surveys (ACS)

4.1 Overview


Tables 4.1 - 4.5 summarize the capabilities of the SIs. For some applications more than one instrument can accomplish a given task, but not necessarily with equal quality or speed. Note that there may be small differences between the numbers quoted here and those quoted in the HST Instrument Handbooks. In such cases the Handbook numbers take precedence.


Table 4.1: HST Instrument Capabilities: Direct Imaging1
SI
Field of View [arcsec]
Projected Pixel Spacing on Sky [arcsec]
Wavelength Range [Å]
Magnitude Limit2
ACS/WFC3
ACS/HRC
ACS/SBC
202 x 202
29 x 26
35 x 31
~0.05
~0.027
~0.032
3700-11,000
2000-11,000
1150-1700
28.8
28.5
24.3
NICMOS/NIC1
NICMOS/NIC2
NICMOS/NIC3
11 x 11
19 x 19
51 x 51
0.043
0.076
0.20
8000-19,000
8000-25,000
8000-25,000
23.2
24.7
25.6
STIS/CCD
STIS/NUV
STIS/FUV
52 x 52
25 x 25
25 x 25
0.05
0.024
0.024
2500-11,000
1650-3100
1150-1700
28.0
24.6
24.0
WFPC24
150 x 150
35 x 35
0.10
0.0455
1150-11,000
1150-11,000
27.5
27.8

Table 4.2: HST Instrument Capabilities: Slit Spectroscopy
SI
Projected Aperture Size
Resolving
Power5
Wavelength
Range [Å]
Magnitude
Limit2
STIS6
52" x (0.05-2)" [optical]
(25-28)" x (0.05-2)"
[UV first order]

(0.1-0.2)" x (0.025-0.2)"
[UV echelle]

25" x (0.05-2)"
[NUV prism]
First order:
~8000
~700
Echelles:
~100,000
~30,000
Prism:
~150

1150-10,300
1150-10,300

1150-3100
1150-3100

1150-3100

14.9, 15.4, 20.1
16.6, 19.2, 21.9

10.9, 12.4
11.6, 14.3

19.8

Table 4.3: HST Instrument Capabilities: Slitless Spectroscopy
SI
Field of View [arcsec]
Projected Pixel Spacing on Sky [arcsec]
Resolving Power
Wavelength Range [Å]
Magnitude Limit2
ACS/WFC grism G800L
202 x 202
~0.05
~100
5500-11000
24.9
ACS/HRC grism G800L
29 x 26
~0.029
~140
5500-11000
24.1
ACS/HRC prism PR200L
29 x 26
~0.027
~100
2000-4000
23.8
ACS/SBC prism PR130L
35 x 31
~0.035
~100
1250-1800
20.7
ACS/SBC
prism PR110L
35 x 31
~0.035
~100
1150-1800
19.7
NICMOS7
51 x 51
0.2
200
8000-25,000
21.6,21.1,18.0
STIS8
52 x 52
25 x 25
0.05
0.024
~700-8000
~150-8000
2000-10,300
1150-3100
See slit spectroscopy above
WFPC28
10 x 10
0.1
~100
3700-9800
25

Table 4.4: HST Instrument Capabilities: Positional Astrometry
SI
Field of View
Precision
(per observation)
Wavelength Range (Å)
Magnitude
FGS1R
69 square arcmin
~1 mas
4700-7100
<16.7

Table 4.5: HST Instrument Capabilities: Binary Star Resolution and Measurements
SI
Field of View
Minimum
Separation [mas]
Accuracy [mas]
Delta Magnitude (max)
Primary Star Magnitude
FGS1R
aperture center
5" x 5" IFOV
8
10
15
20
30
1
1
1
1
1
0.6
1.0
1.0
2.5
4.0
<14.5
<14.5
<16.6
<16.3
<15.0
Notes to Tables 4.1 - 4.5
1 WFPC2, ACS, and NICMOS have polarimetric imaging capabilities. STIS, ACS, and NICMOS have coronagraphic capabilities.
2 Limiting V magnitude for an unreddened A0 V star in order to achieve a signal-to-noise ratio of 5 in a CR-SPLIT (when appropriate) exposure time of 1 hour assuming low-background conditions (LOW; see Section 5.5.1). The limiting magnitude for imaging in the visual is strongly affected by the sky background; under normal observing conditions, the limiting magnitude can be about 0.5 brighter than listed here. Please note that low-sky conditions limit flexibility in scheduling and are not compatible with observing in the CVZ. Single entries refer to wavelengths near the center of the indicated wavelength range. STIS direct imaging entries assume use of a clear filter for the CCD and the quartz filter for the UV (for sky suppression). For STIS spectroscopy to achieve the specified signal-to-noise ratio per wavelength pixel with a 0.5" slit (0.2" for the echelles), multiple values are given corresponding to 1300, 2800 and 6000Å, respectively (if in range). The ACS/WFC, ACS/HRC entries in Table 4.1 refer to an optimum SNR of 5 and assume respectively the filters F606W, F606W and F125LP, the default point source region for the camera selected (0.2 arcsec for WFC and HRC, 0.5 arcsec for SBC), a Bruzual synthetic stellar spectrum, CR SPLIT= 2, and a low background (Zodiacal=low, Earthshine=Average, Airglow=Low). ACS spectroscopy assumes wavelengths of 1300Å (PR130L), 3000Å (PR200L), and 6800Å (G800L). The WFPC2 entries assume F606W. WFPC2 Charge Transfer Efficiency (CTE) losses are negligible for this filter, due to the significant sky background accumulated over 3600 sec in F606W. However, note that WFPC2 images of faint point sources with little sky background can experience significant CTE losses; please see the WFPC2 Instrument Handbook for details. The ACS/SBC entry in Table 4.1 assumes filter F125LP. For NICMOS imaging, we assume filter F160W with a detector temperature of 77.1 K; the limiting H magnitude, in the Vega system, is given for a point source to reach S/N=5 in the brightest pixel and 1 hr exposure. Please see the NICMOS Instrument Handbook, Chapter 9, for details.
3 With ramp filters, the FOV is smaller for the ACS/WFC. Please see the ACS Instrument Handbook for details.
4 The WFPC2 has four CCD chips that are exposed simultaneously. Three are "wide-field" chips, each covering a 75x 75" field and arranged in an "L" shape, and the fourth is a "planetary" chip covering a 35x 35" field.
5 The resolving power is lambda/resolution; for STIS it is /2 where is the dispersion scale in Angstroms/pixel.
6 The 25" or 28" first order slits are for the MAMA detectors, the 52" slit is for the CCD. The R ~150 entry for the prism on the NUV-MAMA is given for 2300Å. More accurate and up to date values for spectroscopic limiting magnitudes can be found in the STIS Instrument Handbook.
7 NICMOS has three grisms (G096,G141,G206) for use in NIC3. We assume a detector temperature of 77.1 K and "average" zodiacal light; the limiting Vega system H magnitude for spectroscopy is given for a point source to reach S/N=5 in 1 hr exposure.
8 All STIS modes can be operated in a slitless manner by replacing the slit by a clear aperture or filter. WFPC2 has a capability of obtaining low-resolution spectra by placing a target successively at various locations in the WFPC2 linear ramp filter. STIS also has a prism for use in the UV.

4.1.1 Instrument Comparison

Observers often face the choice of deciding which HST instrument is best-suited for a particular observation. In some cases, the choice is limited to one instrument, but in many situations the proposer must decide from among several possibilities. Instrument choices for imaging observations currently include ACS, NICMOS, STIS and WFPC2. Spectroscopic observations are currently limited to STIS, except for the slitless capabilities of other instruments listed in Table 4.3:. Some general considerations follow, and further details can be found in the individual instrument handbooks.

The following tables provide some basic recommendations that may be of use in deciding which HST instrument to use in Cycle 13. Table 4.6 summarizes typical decisions that are frequently made when choosing an HST instrument for spectroscopic observations. Table 4.7 lists typical decisions that are often made for imaging observations. All recommendations should be considered general in nature and are meant to provide high-level guidance to observers. The ultimate choice of instrument for a particular observation may depend upon a variety of competing factors and is left as a decision to be made by the proposer.


Table 4.6: Spectroscopy Decisions
Type of Observation
Recommended Instrument
Spectroscopy at R>100
STIS
slitless spectroscopy
ACS (R~100), STIS (R > 100), or NICMOS grism (R~200)

Table 4.7: Imaging Decisions
Type of Observation
Recommended Instrument
Comment
Ultraviolet Observations
< 2000Å
ACS/SBC
Larger FOV, better sensitivity than STIS
> 2000Å
WFPC2 or ACS/HRC or STIS/NUV
WFPC2 has the largest FOV
ACS/HRC has the highest throughput
STIS/NUV has no readout noise
Optical Observations
Broadband > 4000Å
ACS/WFC
Wide field, high throughput
Narrowband > 4000Å
WFPC2 or ACS
WFPC2 has more filter choices
ACS has a few narrow filters
ramp filters are available for smaller areas (both instruments)
High resolution
ACS/HRC
Best sampled PSF
Coronagraphy
ACS/HRC, STIS

Infrared Observations
Wavelength > 1 micron
NICMOS
Only instrument available
Coronagraphy
NICMOS

4.1.2 Future Instruments

Two new science instruments, the Cosmic Origins Spectrograph (COS) and the Wide Field Camera 3 (WFC3), will be installed in HST during SM4. Instrument capabilities are outlined in the COS and WFC3 mini-handbooks released with the Cycle 13 Call for Proposals. Highlights of these instruments are described below.

The Cosmic Origins Spectrograph (COS) is designed to perform high sensitivity, moderate-resolution (R ~ 16,000-24,000) and low-resolution (R ~2000-3000) spectroscopy of astronomical objects in the 1150-3200Å spectral region. COS will significantly enhance the spectroscopic capabilities of HST at ultraviolet wavelengths, and will provide observers with unparalleled opportunities for observing faint sources of ultraviolet light. Wavelength coverage at far-UV wavelengths (1150-2050Å) is expected to be ~300-820Å per exposure, depending upon the spectroscopic mode chosen. Far-UV light is recorded by a crossed delay-line microchannel plate (MCP) detector. Wavelength coverage at near-UV wavelengths (1700-3200Å) is expected to be ~100-800Å per exposure, depending upon the spectroscopic mode chosen. Near-UV light is recorded by a multi-anode microchannel array (MAMA) detector similar to the MAMA detectors used on STIS. COS has two circular science apertures that are 2.5 arcseconds in diameter. Limited COS imaging capabilities are available only at near-UV wavelengths. COS is not meant to be a replacement for STIS, which will remain in HST after the servicing mission and will be available to the community. Rather, COS will complement and extend existing HST spectroscopic capabilities. The high- sensitivity ultraviolet spectroscopy enabled by COS will allow users to investigate fundamental astrophysical topics such as the ionization and baryonic content of the intergalactic medium, the origin of large scale structure in the Universe, the chemical and dynamical evolution of galaxies, and the properties of stars and planetary systems.

The Wide Field Camera 3 (WFC3) is a panchromatic, wide-field, high-throughput imaging camera that will replace WFPC2 in its radial bay during SM4. It features two independent channels - the UVIS channel, sensitive at ultraviolet and optical wavelengths (2000-10,000Å) with a FOV of 2.7 x 2.7 arcmin and a scale of 0.04 arcsec/pixel, and the IR channel, sensitive at near infrared wavelengths (8500Å - 1.7 microns) with a FOV of 2.2 x 2.2 arcmin and a scale of 0.13 arcsec/pixel. The instrument's extended wavelength range, angular resolution, and large field-of-view, along with high sensitivity and a wide selection of spectral elements (62 filters and 1 grism in the UVIS channel; 14 filters and 2 grisms in the IR channel), will provide users with a vast array of options for imaging and low-dispersion, slitless spectroscopy, making WFC3 one of the most versatile instruments onboard HST.

WFC3 will complement and extend the capabilities of ACS and NICMOS, and will also provide a high degree of redundancy with these instruments to help secure HST's unique imaging capabilities until the end of its mission. WFC3 is a facility instrument that will allow users to carry out a variety of key scientific investigations, including searches for galaxies at redshift up to z~10, the study of the physics of star formation in distant and nearby galaxies, accurate measures of the baryonic mass by detecting stars at the limit of the hydrogen-burning sequence and extra-solar Jupiter-like planets, and the study of planetary objects in the Solar System.


Chapter 4:  Cycle 13 Scientific Instruments  4.2 Advanced Camera for Surveys (ACS)
Space Telescope Science Institute
http://www.stsci.edu
Voice: (410) 338-1082
help@stsci.edu