STScI

Cosmic Origins Spectrograph Instrument Handbook for Cycle 17

TOC PREV NEXT INDEX PDF

10.2 Count Rate, Sensitivity, and S/N


10.2.1 Centering accuracy and photometric precision

A complete theoretical discussion of the exposure time as a function of instrument sensitivity and signal-to-noise ratio is given in the Chapter 6 of the STIS Instrument Handbook and will not be repeated here. However, COS has several characteristics which simplify the signal-to-noise calculations.

Both COS detectors are photon counters, which means that they have zero read noise. COS is optimized for point sources, and in this case the signal-to-noise ratio is given by:

where:

C = the signal from the astronomical source, in counts sec-1

t = the integration time, in sec

Npix = the total number of detector pixels integrated to achieve C

Bsky = the sky background, in counts sec-1 pixel-1

Bdet = the detector dark count rate, in counts sec-1 pixel-1

With no detector read noise, the signal-to-noise ratio is proportional to the square root of the exposure time whether the target is bright or faint compared to the backgrounds and dark.


Space Telescope Science Institute
http://www.stsci.edu
Voice: (410) 338-1082
help@stsci.edu
TOC PREV NEXT INDEX PDF