| Cosmic Origins Spectrograph Instrument Handbook for Cycle 17 | ||||
|
|
10.2 Count Rate, Sensitivity, and S/N
10.2.1 Centering accuracy and photometric precision
A complete theoretical discussion of the exposure time as a function of instrument sensitivity and signal-to-noise ratio is given in the Chapter 6 of the STIS Instrument Handbook and will not be repeated here. However, COS has several characteristics which simplify the signal-to-noise calculations.
Both COS detectors are photon counters, which means that they have zero read noise. COS is optimized for point sources, and in this case the signal-to-noise ratio is given by:
C = the signal from the astronomical source, in counts sec-1
t = the integration time, in sec
Npix = the total number of detector pixels integrated to achieve C
Bsky = the sky background, in counts sec-1 pixel-1
Bdet = the detector dark count rate, in counts sec-1 pixel-1
With no detector read noise, the signal-to-noise ratio is proportional to the square root of the exposure time whether the target is bright or faint compared to the backgrounds and dark.
|
Space Telescope Science Institute http://www.stsci.edu Voice: (410) 338-1082 help@stsci.edu |